  "id": "instruments:latnex-hf-b8g",
  "collection": "instruments",
  "slug": "latnex-hf-b8g",
  "name": "LATNEX HF-B8G",
  "summary": "Isotropic tri-axis RF field-strength meter covering 10 MHz to 8 GHz over 38 mV/m to 20 V/m, with per-axis display, max and average readings, a calibration certificate in the box and manual storage of 200 data sets.",
  "status": "verified",
  "lastVerified": "2026-08-12",
  "data": {
    "aliases": [
      "HF-B8G",
      "HFB8G",
      "LATNEX 3-axis RF meter"
    ],
    "manufacturer": "LATNEX",
    "model": "HF-B8G",
    "instrumentClass": "broadband-rf-meter",
    "kind": "broadband",
    "measures": [
      "electric-field-strength",
      "power-density"
    ],
    "bands": [
      {
        "quantity": "electric-field-strength",
        "minHz": 10000000,
        "maxHz": 8000000000,
        "label": "10 MHz – 8 GHz"
      },
      {
        "quantity": "power-density",
        "minHz": 10000000,
        "maxHz": 8000000000,
        "label": "10 MHz – 8 GHz, the same sensor read out as power density"
      }
    ],
    "axes": "triple",
    "ranges": [
      {
        "mode": "RF field strength, isotropic",
        "range": "38 mV/m – 20 V/m",
        "resolution": "0.1 mV/m; 0.001 µW/m²"
      }
    ],
    "calibration": "Supplied with a calibration certificate according to the manufacturer's listing, plus a user calibration factor settable from 0.10 to 9.99 — a correction factor, not a traceable calibration.",
    "peakHold": true,
    "logging": true,
    "suitedTo": [
      "Isotropic RF surveys where orientation should not change the reading: all three axes are summed and each can also be read individually",
      "Wide low-end coverage down to 10 MHz, which takes in shortwave and FM broadcast that most consumer RF meters miss",
      "Recording a set of survey positions: 200 manual data sets stored on the meter"
    ],
    "cannotDo": [
      "No low-frequency magnetic or electric field measurement",
      "Nothing above 8 GHz",
      "The 38 mV/m floor is around 3.8 µW/m², so it cannot resolve the very low levels precautionary sleeping-area targets are written at",
      "No spectrum or frequency counter, so a reading cannot be attributed to a service",
      "A supplied certificate is the manufacturer's own; it is not an ISO/IEC 17025 accredited calibration unless the certificate says so"
    ],
    "marketStatus": "current",
    "availability": {
      "vendor": "EMR Shielding Solutions",
      "url": "https://www.emrss.com/products/hf-b8g-rf-meter",
      "relationship": "owned"
    },
    "related": [
      "meters:broadband-rf-meter",
      "protocols:rf-room-survey",
      "quantities:electric-field-strength"
    ]
  },
  "sources": [
    {
      "id": "latnex-hf-b8g",
      "title": "LATNEX HF-B8G professional RF meter — specifications",
      "publisher": "LATNEX",
      "url": "https://www.latnex.com/products/hf-b8g",
      "kind": "manufacturer",
      "retrieved": "2026-08-12"
    }
  ],
  "links": {
    "self": "https://emfbase.com/api/instruments/latnex-hf-b8g",
    "html": "https://emfbase.com/instruments/latnex-hf-b8g",
    "markdown": "https://emfbase.com/instruments/latnex-hf-b8g.md",
    "collection": "https://emfbase.com/api/instruments"
  }
}