EMFBase

Safe and Sound Pro II

Single-axis broadband RF meter with a stated ±6 dB response from 400 MHz to 7.2 GHz, linear display from 0.1 to 1,000,000 µW/m², sub-5 µs pulse response and peak, max and average readings.

Broadband meterSafe Living Technologies Inc.single-axisCurrent model

Last verified 12 August 2026.

Also searched as: SS Pro II, Safe and Sound Pro 2, Safe Living Technologies Pro II

Manufacturer
Safe Living Technologies Inc.
Model
Safe and Sound Pro II
Type
Broadband meter
Instrument class
Broadband RF power-density meter
Quantities measured
power density, electric field strength
Axes
single
Peak hold
Yes
On-device logging
No
Declared accuracy
±6 dB over 400 MHz – 7.2 GHz, temperature compensated −20 to +60 °CManufacturer figure, not an independent verification.
Market status
Currently sold

Frequency coverage by mode

One band per measuring mode, as the manufacturer states it.

power density
200 MHz – 8 GHz full response; ±6 dB from 400 MHz to 7.2 GHz
electric field strength
200 MHz – 8 GHz, the same reading switchable to V/m

Display range by mode

RF power density, linear
0.1 – 1,000,000 µW/m²
RF power density, reduced tolerance
0.005 – 3,180,000 µW/m² (0.001 – 34.6 V/m)

Calibration

Manufacturer-stated factory calibration. Ask for the current calibration statement before a reading has to stand up formally; the manual makes no accredited-laboratory claim.

Suited to

  • Room and sleeping-area RF surveys where a total exposure figure is wanted
  • Detecting short pulsed transmissions — the manual claims response to pulses under 5 µs, which covers Wi-Fi and 5G bursts
  • Identifying a source by ear through the built-in speaker before deciding what to change

What it cannot do

  • No power-frequency magnetic or electric field measurement — wiring, transformers and power lines need a separate instrument
  • Nothing above 8 GHz, so mmWave 5G at 24 GHz and above is invisible to it
  • No spectrum display and no frequency counter: the reading is a total across the band, so it cannot say which service produced it
  • Single-axis, so the reading depends on orientation; the meter has to be rotated to find the maximum
  • No on-device logging, so a 24-hour record needs a different instrument or an external recorder

What these figures are

Every number on this page is the manufacturer's own declared specification, reproduced so that models can be compared field by field. Nothing here is an independent test of the instrument, a calibration certificate, or a statement that a reading taken with it can be compared with a regulatory limit — that depends on the detector, the averaging and, where it matters legally, an accredited calibration.

Related records

  • Broadband RF power-density meter

    The general-purpose instrument for Wi-Fi, cellular, 5G, smart meters and cordless phones: reports total power density across a wide band in µW/m² or mW/m², with peak-hold and average modes.

    Instruments

  • RF room survey

    How to measure radiofrequency exposure in a room so the result is reproducible: peak and average, at body positions, with the dominant direction identified.

    Measurement protocols

  • Power density

    Radiofrequency energy flow per unit area — µW/m² on consumer RF meters, mW/cm² in FCC rules, W/m² in ICNIRP guidelines. 1 mW/cm² = 10 W/m² = 10,000,000 µW/m².

    Quantities & units

Sources

  1. slt-safe-sound-pro-ii Safe and Sound Pro II operation manual and specifications Safe Living Technologies Inc.. manufacturer, link checked 2026-08-12.

Same record, other formats