ELF magnetic field survey
A source-finding walkthrough plus a 24-hour log, because power-frequency magnetic fields track electrical load and a spot reading answers almost nothing.
magnetic flux density
Last verified 12 August 2026.
Also searched as: how to measure milligauss, gaussmeter survey
- Quantity measured
- magnetic flux density
Goal
Find the source of any elevated 50/60 Hz magnetic field and quantify exposure at sleeping and working positions over a full day.
Equipment
- ELF gaussmeter (power-frequency magnetic field)
- Logging EMF instrument
Procedure
- Record meter model, axis configuration and frequency response.
- Walk the perimeter of each room with the meter at waist height, noting every local maximum and its apparent source.
- Measure at 0.3 m, 1 m and 3 m from each candidate source to confirm the inverse fall-off that identifies a compact source.
- Check the service panel, meter base, and both sides of every wall that backs onto one.
- If readings are elevated everywhere and do not fall off with distance, suspect net current on the neutral or a shared neutral, and involve an electrician.
- Log continuously for at least 24 hours at the pillow position, sampling at least once per second where the instrument allows.
- Report the mean, the night-time mean and the 95th percentile, not just the maximum.
Pitfalls
- Spot-reading at one time of day, when line and household loading vary by several times
- Using a single-axis meter and rotating it inconsistently
- Assuming shielding is the fix; magnetic fields pass through nearly all common materials
- Missing a wiring fault, which is the one cause that raises the whole building and cannot be solved with distance
- Using a 50/60 Hz-only meter near inverters, EV chargers or induction hobs, where the energy is in the kilohertz range
Interpret results against
- Building Biology SBM-2024 AC magnetic field bands
- WHO/IARC ELF magnetic field association threshold (0.3–0.4 µT)
- ICNIRP 2010 general public magnetic field reference level at 50/60 Hz
- Swiss installation limit value for power lines
Related records
- Electrical service panel and meter base
The highest-current point in a building: readings in the tens of milligauss are normal within a metre, falling to background within two to three metres.
Sources & emitters
- ELF gaussmeter (power-frequency magnetic field)
Measures 50/60 Hz magnetic flux density in milligauss or microtesla — the instrument for wiring faults, service panels, transformers and power lines.
Instruments
Sources
- sbm-2024 Building Biology Evaluation Guidelines for Sleeping Areas (SBM-2024) — Institut für Baubiologie + Nachhaltigkeit (IBN), 2024. ngo, link checked 2026-08-12.
- niehs-emf Electric and Magnetic Fields — National Institute of Environmental Health Sciences. government, link checked 2026-08-12.
- who-elf Electromagnetic fields and public health: exposure to extremely low frequency fields — World Health Organization, 2007. government, link checked 2026-08-12.