Multi-field meter combining RF power density from 100 MHz to 8 GHz, a 50 Hz – 10 kHz magnetic mode, a 50 Hz – 50 kHz electric-field mode, a 100 MHz – 4.2 GHz frequency counter and logging to 1,000 samples.

Multi-field meterCORNET Microsystems Inc.single-axisCurrent model

Last verified 12 August 2026.

Also searched as: ED88T Plus 5G2, Cornet ED88TPlus5G, ED88T

Manufacturer
CORNET Microsystems Inc.
Model
ED88TPlus5G2
Type
Multi-field meter
Instrument class
Broadband RF power-density meter
Quantities measured
power density, electric field strength, magnetic flux density, frequency
Axes
single
Peak hold
Yes
On-device logging
Yes
Declared accuracy
Manufacturer-stated error: RF ±3.5 dB, LF magnetic 20%, electric field 25%Manufacturer figure, not an independent verification.
Market status
Currently sold

Frequency coverage by mode

One band per measuring mode, as the manufacturer states it.

power density
RF: 100 MHz – 8 GHz
magnetic flux density
LF magnetic: 50 Hz – 10 kHz
electric field strength
LF electric: 50 Hz – 50 kHz
frequency
Frequency counter: 100 MHz – 4.2 GHz

Display range by mode

RF power density
0.5 µW/m² – 1.8 W/m² (−60 to +5 dBm)
RF field strength
14 mV/m – 26.2 V/m
LF magnetic, 50 Hz – 10 kHz
0.1 – 60 µT (1 – 600 mG)
LF magnetic, 50 Hz – 1 kHz
0.01 – 1 µT (0.1 – 10 mG)
LF electric
10 – 1,000 V/m

Calibration

No calibration certificate is offered in the manual; the meter has a user-settable frequency-band correction factor rather than a traceable calibration.

Suited to

  • A first survey where RF, wiring magnetic fields and electric fields all have to be looked at with one instrument
  • Attributing an RF reading to a service, because the frequency counter reports the dominant carrier up to 4.2 GHz
  • Unattended recording: 1,000 samples down to 0.001 s intervals, exportable over USB

What it cannot do

  • Not a three-axis instrument in any mode — every reading is orientation dependent
  • The frequency counter stops at 4.2 GHz, so a 5 GHz Wi-Fi or 6 GHz reading cannot be identified even though it is measured
  • Above 8 GHz is out of specification, and the '10 GHz' figure in the maker's comparison table is a coverage claim rather than a specified measurement range
  • The stated LF magnetic and electric tolerances of 20% and 25% are far wider than a metrology-grade gaussmeter, so it will not settle a marginal comparison against a limit
  • No accredited calibration, so its numbers do not carry evidential weight in a formal complaint

What these figures are

Every number on this page is the manufacturer's own declared specification, reproduced so that models can be compared field by field. Nothing here is an independent test of the instrument, a calibration certificate, or a statement that a reading taken with it can be compared with a regulatory limit — that depends on the detector, the averaging and, where it matters legally, an accredited calibration.

Where to buy

EMR Shielding Solutions

EMR Shielding Solutions is a related company of WebKMS, which publishes EMFBase, and sells this product. The link is a stockist listing only: every figure on this page comes from the manufacturer's own documentation, and no specification, limitation or citation depends on it.

Related records

  • Broadband RF power-density meter

    The general-purpose instrument for Wi-Fi, cellular, 5G, smart meters and cordless phones: reports total power density across a wide band in µW/m² or mW/m², with peak-hold and average modes.

    Instruments

  • ELF gaussmeter (power-frequency magnetic field)

    Measures 50/60 Hz magnetic flux density in milligauss or microtesla — the instrument for wiring faults, service panels, transformers and power lines.

    Instruments

  • Logging EMF instrument

    Any of the above with unattended recording: required whenever the assessment depends on an average, a night-time value or a 95th percentile rather than a spot reading.

    Instruments

  • RF room survey

    How to measure radiofrequency exposure in a room so the result is reproducible: peak and average, at body positions, with the dominant direction identified.

    Measurement protocols

Sources

  1. cornet-ed88tplus5g2-manual CORNET ED88TPlus5G2 electrosmog meter user's manual v2.0 CORNET Microsystems Inc., 2023. manufacturer, link checked 2026-08-12.
  2. cornet-selection-guide CORNET electrosmog meter selection guide — model comparison table CORNET Microsystems Inc.. manufacturer, link checked 2026-08-12.

Same record, other formats